Home Page
Thermocouple: K type 1/4"OD x 12" L with Exposed Head and Male Plug for Faster Response - EQ-TC-K-Cali-12S-DT-LD
Important Notice
This is our archived version of MTI old website. Price, inventory, and technical information IS NOT accurate. Please do not use them as factors for purchasing decisions. Please navigate our new web MTI's New Website. Thank you for your understanding! |
|
|
In stock
Item Number: TCK12SDT
|
Features:
Specifications

|
1/4" thermocouple can be inserted into processing tube via flange
and calibrating tube furnace temperature under vacuum *( click picture to order the flange)
|
|
Application Notes
|
- If necessary, please check out K type thermocouple extension wire with mini male plug.
- There are 2 polarities for K type thermocouple: positive and negative. Please check polarity of K type thermocouple before use. Wrong polarity may cause temperature measuring error.
- It is also easy to test the polarity of K type thermocouples. The negative wire is MORE magnetic than the positive wire. Just put a magnet up to each wire. One will be more magnetic than the other.
|
|
Warning
|
- Chromel/Alumel is vulnerable to sulfidic corrosion which can result in a thermocouple failure. Apply a thin layer of Zirconia coating will protect the TC from corrosion and thus extend its service life. Re-apply the coating if there is a sign of layer flaking/peeling after the process. Please click the picture in below to order the product.
|
|
|
MTI sponsorships: MTI Sponsors Thermoelectrics Workshop
MTI-UCSD Battery Fabrication Lab
MTI Sponsors the Postdoctoral Awards